Development of an optical pH measurement system based on colorimetric effect

This paper proposed an optical pH measurement system developed to measure the changes of pH levels based on colorimetric reactions with phenol red reagent. The optical sensing was achieved through the implementation of a pair of light emitting diode (LED) and photodiode. The detection mechanism was...

Full description

Saved in:
Bibliographic Details
Main Authors: Mustafa, Puteri Nur Syahirah Mohamed (Author), Azam, Aiman Shahmi (Author), Sulaiman, Mohd Suhaimi (Author), Omar, Ahmad Fairuz (Author), Rahman, Mohamad Faizal Abd (Author)
Other Authors: Universiti Teknologi MARA Cawangan Pulau Pinang (Contributor)
Format: EJournal Article
Published: Institute of Advanced Engineering and Science, 2021-11-01.
Subjects:
Online Access:Get fulltext
Tags: Add Tag
No Tags, Be the first to tag this record!
LEADER 02392 am a22003373u 4500
001 ijeecs25488_15664
042 |a dc 
100 1 0 |a Mustafa, Puteri Nur Syahirah Mohamed  |e author 
100 1 0 |a Universiti Teknologi MARA Cawangan Pulau Pinang  |e contributor 
700 1 0 |a Azam, Aiman Shahmi  |e author 
700 1 0 |a Sulaiman, Mohd Suhaimi  |e author 
700 1 0 |a Omar, Ahmad Fairuz  |e author 
700 1 0 |a Rahman, Mohamad Faizal Abd  |e author 
245 0 0 |a Development of an optical pH measurement system based on colorimetric effect 
260 |b Institute of Advanced Engineering and Science,   |c 2021-11-01. 
500 |a https://ijeecs.iaescore.com/index.php/IJEECS/article/view/25488 
520 |a This paper proposed an optical pH measurement system developed to measure the changes of pH levels based on colorimetric reactions with phenol red reagent. The optical sensing was achieved through the implementation of a pair of light emitting diode (LED) and photodiode. The detection mechanism was based on different absorbance of light intensity at pH values of 2.36, 8.32, 9.08 and 12.83. The data processing method was carried out using LabVIEW software and interfaced with NI USB DAQ 6008. The measured voltage showed a good correlation in relation to the pH level with R2 equals to 0.9624. This relationship was used as a calibration curve in the final system testing. The final measurement of pH showed a good agreement between the actual and measured values with an error of less than 5%, thus indicating the reliability of the proposed system. 
540 |a Copyright (c) 2021 Institute of Advanced Engineering and Science 
540 |a http://creativecommons.org/licenses/by-nc/4.0 
546 |a eng 
690 |a sensors and instrumentation; optical sensing; water quality; microfluidic 
690 |a Absorbance; Colorimetric; LED based; Optical sensing; pH measurement; 
655 7 |a info:eu-repo/semantics/article  |2 local 
655 7 |a info:eu-repo/semantics/publishedVersion  |2 local 
655 7 |2 local 
786 0 |n Indonesian Journal of Electrical Engineering and Computer Science; Vol 24, No 2: November 2021; 728-735 
786 0 |n 2502-4760 
786 0 |n 2502-4752 
786 0 |n 10.11591/ijeecs.v24.i2 
787 0 |n https://ijeecs.iaescore.com/index.php/IJEECS/article/view/25488/15664 
856 4 1 |u https://ijeecs.iaescore.com/index.php/IJEECS/article/view/25488/15664  |z Get fulltext